Double-quencher probes improve detection sensitivity toward Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) in a reverse-transcription polymerase chain reaction (RT-PCR) assay
Detection Principle
NucleicAcid-PCR based
Target
N1 and N2
Testing Method Category
RT-PCR
Testing Method
2019-nCoV RUO Kit
Testing Method - Additional Info
Double-quencher(YCH-N1) probes improve detection sensitivity comparison with NIID
Reported Performance
Positive percentage agreement NIID-N1: 56%, Positive percentage agreement YCH-N1:61%, Positive percentage agreement YCH-N2: 94%
Sample Size
66 nasopharyngeal swabs
Peer-reviewed
yes