Clinical evaluation of five different automated SARS-CoV-2 serology assays in a cohort of hospitalized COVID-19 patients
Detection Principle
ImmunoAssay-Others
Target
IgG IgM
Testing Method Category
Mixed
Testing Method
Anti-SARS-CoV-2 ELISA (IgG)EUROIMMUN; LIAISON SARS-CoV-2 S1/S2 IgG DiaSorin S; Elecsys Anti-SARS-CoV-2 Roche Diagnostics; WANTAI SARS-CoV-2 Ab ELISA Beijing Wantai Biological Pharmacy Enterprise Co; Atellica IM SARS-CoV-2 Total (COV2T) Siemens Healthcare Gmbh
Testing Method - Additional Info
Comparison of 5 different commercial methods for detecting coronavirus
Reported Performance
Positivity assay Diasorin: 68.4 %, Positivity assay Euroimmun: 70.3 %, Positivity assay Siemens: 73.7 %, Positivity assay Roche: 79.0 %, Positivity assay Wantai: 81.6 %. Specificity Euroimmun: 99,1% Specificity Diasorin: 99.1 %, Specificity Wantai: 99.4 %, Specificity Roche: 99.7 %, Specificity: Siemens 100 %
Sample Size
75 confirmed
Peer-reviewed
yes
The database contains available information from scientific literature that is being updated periodically. Please note that the provided information (as retrieved from analysed papers) is provided only for devices commercially available with CE-IVD mark. Acknowledgements