Analytical and Clinical Evaluation of the Automated Elecsys Anti-SARS-CoV-2 Antibody Assay on the Roche cobas e602 Analyzer

Detection Principle
ImmunoAssay-Antibody
Target
IgA/IgG
Testing Method Category
 
Testing Method
Kit
Testing Method - Additional Info
Elecsys Anti-SARS-CoV-2 immunoassay (Roche Diagnostics) vs Euroimmun IgA/IgG
Reported Performance
Sensitivity: 95% (>14 days post positive PCR), 77.5% (0-13 days post positive PCR); specificity: 100%; Concordance with ELISA: 100%; PPV: 100%; NPV: 98.83%; accuracy: 98.88 (0-13 days post positive PCR), 99.74% (>14 days post positive PCR)
Sample Size
66 samples (method comparison) & 77 samples (cross reactivity) & 78 samples (clinical performance)