Skip to main content
COVID-19 In Vitro Diagnostic Devices and Test Methods Database

Scientific literature on COVID-19 Test Methods and Devices - detail

Comparison of four new commercial serologic assays for determination of SARS-CoV-2 IgG

Detection Principle
ImmunoAssay-Antibody
Target
IgG
Testing Method Category
 
Testing Method
kit
Testing Method - Additional Info
Euroimmun is validated in Germany
Reported Performance
Sensitivity: 86.4% (Euroimmun), 100% (EDI), 86.4% (Mikrogen), 77.3% (Viramed); specificity: 96.2% (Euroimmun), 88.7% (EDI), 100% (Mikrogen), 100% (Viramed)
Sample Size
75 samples
Peer-reviewed
yes

The database contains available information from scientific literature that is being updated periodically. Please note that the provided information (as retrieved from analysed papers) is provided only for devices commercially available with CE-IVD mark. Acknowledgements