Comparison of commercial RT-PCR diagnostic kits for COVID-19

Detection Principle
NucleicAcid-PCR based
Target
E, S, ORF1ab, N, RdRp genes
Testing Method Category
PCR_KIT
Testing Method
7 commercial RT-PCR kits compared
Reported Performance
PCR efficiency: >96% (all tests) ; sensitivity: 13/16 (R-Biopharm AG), 12/16 (BGI, KH Medical, Seegene), 11/16 (CerTest BioTec), 10/16 (Altona Diagnostic, PrimerDesign) ; LOD95%: 3.8 cpu (Altona Diagnostic E- and S-gene), 23 cpu (PrimerDesign), 0.91 cpu (in-house Corman E-gene), 4.3 cpu (BGI and R-Biopharm AG, KH Medical S-gene), 4.8 cpu (CerTest Biotec N-gene, KH Medical ORF1ab/RdRp), 18 cpu (CerTest Biotec Orf1ab/RdRp)
Sample Size
16 patients Covid19 positive, 6 patients other respiratory viral infections
Peer-reviewed
no