Development and validation of direct RT-LAMP for SARS-CoV-2

Detection Principle
NucleicAcid-PCR based
Target
S, RdRP, ORF1ab
Testing Method Category
LAMP
Testing Method
In-house LAMP
Reported Performance
PPA: 97.62% (Sgene), 91.97% (S and RdRP); NPA: 98.72% (Sgene), 100% (S and RdRP)
Sample Size
120 samples
Peer-reviewed
no

The database contains available information from scientific literature that is being updated periodically. Please note that the provided information (as retrieved from analysed papers) is provided only for devices commercially available with CE-IVD mark. Acknowledgements