Analytical Sensitivity and Specificity of Two RT-qPCR Protocols for SARS-CoV-2 Detection Performed in an Automated Workflow.

Detection Principle
NucleicAcid-PCR based
Target
N1, E, RdRP (modified)
Testing Method Category
PCR_KIT
Testing Method
kits (Charité and Centers for Disease Control (CDC)
Reported Performance
LOD: 21 cpr (N1), 141 cpr (E), 350 cpr (RdRP), 33.7 cpr (RdRP modified); PPA: 91.3% (RdRP vs N1), 96% (E vs N1); NPA: 99.6% (RdRP vs N1), 99.8% (E vs N1)
Sample Size
60 samples, 968 samples (accuracy)
Peer-reviewed
yes

The database contains available information from scientific literature that is being updated periodically. Please note that the provided information (as retrieved from analysed papers) is provided only for devices commercially available with CE-IVD mark. Acknowledgements