Validation of the STANDARD Q COVID-19 antigen test in Vojvodina Serbia.

Detection Principle
ImmunoAssay-Antigen
Testing Method Category
LFIA
Testing Method
STANDARD Q COVID-19 Ag Test (SD Biosensor, Gyeonggi-do, South Korea)
Testing Method - Additional Info
Comparison of chromatographic antigen assay with Argene, SARS-COV-2 R-GENE assay (bioMerieux, Marcy-l’Etoile, France) and COVID-19 Genesig Real-Time PCR Kit (Primerdesign Ltd, Chandler’s Ford, UK)
Reported Performance
Sensitivity: 58.1% ; Specificity: 100% ; PPV: 100% ; NPV: 81.1% ; Accuracy: 85% ; remarks: more values are available for sample subsets
Sample Size
120 patients
Peer-reviewed
yes

The database contains available information from scientific literature that is being updated periodically. Please note that the provided information (as retrieved from analysed papers) is provided only for devices commercially available with CE-IVD mark. Acknowledgements