Evaluation of seven commercial SARS-CoV-2 RNA detection kits based on real-time polymerase chain reaction (PCR) in China

Detection Principle
NucleicAcid-PCR based
Target
N, ORF1ab, E genes
Testing Method Category
PCR_KIT
Testing Method
kit A (BGI Biotech Co., Ltd, Wuhan, China), kit B (Outdo Biotech Co., Ltd, Shanghai, China), kit C (Sansure Biotech Inc., Changsha, China), kit D (Perkin Elmer Medical Diagnostic Products, Co., Ltd, Shanghai, China), kit E (Daan Gene Co., Ltd. Of Sun Yat-Sen University, Guangzhou, China), kit F (Jiangsu Bioperfectus Technologies Co., Ltd, Taizhou, China), and kit G (Fosun Long March Medical Science Co., Ltd, Shanghai, China).
Reported Performance
PPV: 100% (all kits) ; sensitivity: 90.48% (A, F), 92.86% (B), 83.33% (C), 97.62% (D), 78.57% (E), 76.19% (G) ; Specificity: 100% (all) ; NPA: 98.04% (A,F), 98.52% (B), 96.62% (C), 99.50% (D), 95.69% (E), 95.24% (G)
Sample Size
242 specimens