COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Antigen Test Kit

Manufactured by BIOTEKE CORPORATION (WUXI) Co. Ltd., China - https://www.bioteke.cn/ 

Device identification number
5853
CE Marking
Yes
HSC common list (RAT)
Yes
Format
Near POC / POC
Physical Support
Cassette
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Anterior nasal swab
Cross-reactivity (pathogens tested)
SARS-CoV-2
Pathogens detected
SARS-CoV-2
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), Other (), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2023-02-13 08:48:31 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
N.A.
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
Detection Principle
N.A.
LOD
1250 TCID50/ml
Calibration
Not evaluated
Analytical Sensitivity
100 %
Analytical Specificity
100 %
Analysis of cross reactivity
Evaluated
False positives
8.65 %
False negatives
0.27 %
Precision
Evaluated
Accuracy
97.91 %
Robustness
Evaluated
Clinical Sensitivity
91.35 %
Clinical Specificity
99.73 %