CHECK UP COVID19-INF A/B-RSV MULTI RAPID TEST KIT
Manufactured by Cesna Biyoteknoloji Araştırma Geliştirme Laboratuvar Sist.İnş.Müh.Dan.San.Tic.Ltd.Şti., Turkey - www.cesnalaboratuvarsistemleri.com
Device identification number
5817
CE Marking
✓Yes
HSC common list (RAT)
×No
Format
Manual, Near POC / POC
Physical Support
Cassette
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Anti-Nuclear Antibody, Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Cytomegalovirus (CMV), Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Hemophilus Influenzae, Hepatitis A Virus (HAV), Hepatitis B Virus (HBV), Hepatitis C Virus (HCV), Herpes Simplex (HSV), Human Immunodeficiency Virus (HIV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mumps Virus (MuV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV, SARS-CoV-2, Varicella Zoster Virus (VZV)
Pathogens detected
Other (-)
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2023-01-09 10:08:59 CET
Comments
In addition, our Covid19 test kit has been published by Common list with RAT number 2696.
We were verified by PEI in 2021.
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
Yes
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
Detection Principle
Colloidal gold
LOD
3200 TCID50/ml (for Covid19)
Calibration
Evaluated
Analytical Sensitivity
100 % (for Covid19)
Analytical Specificity
100 % (for Covid19)
Analysis of cross reactivity
Evaluated
False positives
0.33 % (for Covid19)
False negatives
0.49 % (for Covid19)
Precision
Evaluated
Accuracy
99.6 % (for Covid19)
Robustness
Evaluated
Clinical Sensitivity
99.3 % (for Covid19)
Clinical Specificity
99.7 % (for Covid19)