COVID-19 & Influenza A/B& RSV Antigen Nasal Test Kit

Manufactured by Assure Tech (Hangzhou) Co. Ltd., China - www.diareagent.com 

Device identification number
5803
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Human Metapneumovirus (HMPV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Rhinovirus A, Rhinovirus B
Pathogens detected
Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza B, Influenza B Victoria, Influenza B Yamagata, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, SARS-CoV
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-12-22 08:48:45 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
Detection Principle
Other (chromatographic immunoassay)
LOD
251 TCID50/ml
Calibration
Evaluated
Analytical Sensitivity
97 %
Analytical Specificity
99.3 %
Analysis of cross reactivity
Evaluated
False positives
0.7 % (A total of 450 SARS-CoV-2 negative samples were tested, 3 samples were positive by this test.)
False negatives
3 % (A total of 101 SARS-CoV-2 positive samples were tested, 3 samples were negative by this test.)
Precision
Evaluated
Accuracy
98.9 %
Robustness
Evaluated
Clinical Sensitivity
97 %
Clinical Specificity
99.3 %

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements