V-Chek 2019-nCoV Ag Rapid Test Kit (Immunochromatography)

Manufactured by Guangzhou Decheng Biotechnology Co. Ltd., China - www.dochekbio.com 

Device identification number
5762
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual
Physical Support
Cassette
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus 3, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Epstein-Barr Virus (EBV), Human Metapneumovirus (HMPV), Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type 2, Respiratory Syncytial V (RSV), Rhinovirus
Pathogens detected
Adenovirus 3, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Epstein-Barr Virus (EBV), Human Metapneumovirus (HMPV), Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type 2, Respiratory Syncytial V (RSV), Rhinovirus
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Information on antigen epitope targeted by the antibodies used in the design of the device
74-105aa
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.429 (Epsilon), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2022-11-21 09:01:32 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
Yes
Reader Required
No
Subcategory
N.A.
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
Detection Principle
Colloidal gold
LOD
100 TCID50/ml
Calibration
Not evaluated
Analytical Sensitivity
%
Analytical Specificity
%
Analysis of cross reactivity
Evaluated
False positives
0.58 %
False negatives
3.45 %
Precision
Evaluated
Accuracy
98.39 %
Robustness
Evaluated
Clinical Sensitivity
96.55 %
Clinical Specificity
99.42 %