SARS-CoV-2 & Influenza A/B Antigen Combo Test Kit(Colloidal Gold Chromatographic Immunoassay)
Manufactured by Shenzhen Microprofit Biotech Co. Ltd., China - www.microprofit-bio.com
Device identification number
5751
CE Marking
✓Yes
HSC common list (RAT)
✓Yes
Format
Manual
Physical Support
Cassette, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab, Nasopharyngeal swab, Oropharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Cytomegalovirus (CMV), Epstein-Barr Virus (EBV), Hemophilus Influenzae, Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, SARS-CoV, SARS-CoV-2
Pathogens detected
Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, SARS-CoV-2
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.37 (Lambda), P.1 (Gamma), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-11-03 08:59:31 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
Detection Principle
Colloidal gold
LOD
49 TCID50/ml
Calibration
Evaluated
Analytical Sensitivity
100 %
Analytical Specificity
100 %
Analysis of cross reactivity
Evaluated
False positives
%
False negatives
6.79 %
Precision
Evaluated
Accuracy
96.26 %
Robustness
Evaluated
Clinical Sensitivity
93.21 %
Clinical Specificity
100 %
The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements