COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Antigen Test Kit (Colloidal Gold) (Self-Testing)

Manufactured by Genrui Biotech Inc., China - www.genrui-bio.com 

Device identification number
4159
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual, Other ((Self-Testing))
Physical Support
Cassette, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Enterovirus A71 (EV-A71), Hemophilus Influenzae, Human Metapneumovirus (HMPV), Influenza A, Influenza B, MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus
Pathogens detected
SARS-CoV-2
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), Other ((B.1.618, B.1.640.2, AY.1, AY.4.2, BA.1, BA.1.1, BA.2, BA.3, XE, BA.4, BA.5, BA.2.12.1, BA.2.12, BA.2.75, BF.7)), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-10-14 10:51:22 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
Yes
Reader Required
N.A.
Subcategory
N.A.
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich
Detection Principle
Colloidal gold
LOD
180 TCID50/ml
Calibration
Not evaluated
Analytical Sensitivity
100 %
Analytical Specificity
100 %
Analysis of cross reactivity
Evaluated
False positives
0 %
False negatives
1.87 %
Precision
Evaluated
Accuracy
99.03 %
Robustness
Evaluated
Clinical Sensitivity
98.13 %
Clinical Specificity
100 %

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements