Rapid nCoV & Flu A/B Ag Test Kit (Colloidal Gold)

Manufactured by Shenzhen Huian Biosci Technology Co. Ltd., China - www.huiantech.com 

Device identification number
4098
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual, Near POC / POC
Physical Support
Cassette, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab, Nasopharyngeal swab, Oropharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Anti-Nuclear Antibody, Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Cytomegalovirus (CMV), Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Hemophilus Influenzae, Hepatitis A Virus (HAV), Hepatitis B Virus (HBV), Hepatitis C Virus (HCV), Herpes Simplex (HSV), Human Immunodeficiency Virus (HIV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mumps Virus (MuV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Other (), Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV, SARS-CoV-2, Varicella Zoster Virus (VZV)
Pathogens detected
Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza B, Influenza B Victoria, Influenza B Yamagata, SARS-CoV-2
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.621 (Mu), C.37 (Lambda), P.1 (Gamma), P.2 (Zeta)
Commercial Status
Commercialised
Last Update
2022-09-29 09:01:44 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
Detection Principle
Colloidal gold
LOD
150 TCID50/ml (SARS-CoV-2)
Calibration
Not evaluated
Analytical Sensitivity
94.6 % (SARS-CoV-2)
Analytical Specificity
99.1 % (SARS-CoV-2)
Analysis of cross reactivity
Evaluated
False positives
0.9 % (SARS-CoV-2)
False negatives
5.4 % (SARS-CoV-2)
Precision
Evaluated
Accuracy
97.6 % (SARS-CoV-2)
Robustness
Evaluated
Clinical Sensitivity
94.6 % (SARS-CoV-2)
Clinical Specificity
99.1 % (SARS-CoV-2)