Flowflex SARS-CoV-2 Antigen Saliva Rapid Test Lollipop (Self-Testing)

Manufactured by Acon Biotech (Hangzhou) Co. Ltd., China - https://www.aconbio.com/ 

Device identification number
4051
CE Marking
Yes
HSC common list (RAT)
×No
Format
Other (Self-Testing)
Physical Support
Cassette, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Saliva
Cross-reactivity (pathogens tested)
SARS-CoV
Pathogens detected
SARS-CoV, SARS-CoV-2
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), Other (B.1.618, B.1.640.2, AY.1, AY.4.2, BA.1, BA.1.1, BA.2, BA.3,XE,BA.4,BA.5, BA.2.12.1,BA.2.12, BA.2.75), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-09-20 01:32:38 CET
Comments
N.A
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
Yes
Reader Required
No
Subcategory
N.A.
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
Detection Principle
N.A.
LOD
320 TCID50/ml
Calibration
Not evaluated
Analytical Sensitivity
100 %
Analytical Specificity
100 %
Analysis of cross reactivity
Evaluated
False positives
0 %
False negatives
18.2 %
Precision
Evaluated
Accuracy
96.4 %
Robustness
Evaluated
Clinical Sensitivity
81.8 %
Clinical Specificity
100 %