Red Cell COVID-19 Self Test
Manufactured by Red Cell Biyoteknoloji A.Ş., Turkey - www.redcell.com.tr
Device identification number
3972
CE Marking
✓Yes
HSC common list (RAT)
×No
Format
Manual, Near POC / POC
Physical Support
Cassette, Lateral flow, Latex
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Anterior nasal swab, Mid-turbinates swab, Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Herpes Simplex (HSV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mumps Virus (MuV), Mycobacterium Tuberculosis, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 3, Respiratory Syncytial V (RSV), SARS-CoV
Pathogens detected
Other (No Cross Reactivity has been detected)
Information on antigen epitope targeted by the antibodies used in the design of the device
N terminal (48-113)
NTASWFTALTQHGKEDLKFPRGQGVPINTNSSPDDQIGYYRRATRRIRGGDGKMKDLSPRWYFYYL
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
Yes
Reader Required
No
Subcategory
Other ()
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
Detection Principle
Other ()
LOD
33 PFU/mL
Analytical Sensitivity
100 %
Analytical Specificity
100 %
False positives
0 N.A.
False negatives
0.025 ng/mL
Accuracy
97.9 %
Clinical Sensitivity
96.3 %
Clinical Specificity
99.5 %