COVID-19 & Influenza A+B Antigen Combo Rapid Test (For Self-Testing)

Manufactured by Safecare Biotech (Hangzhou) Co. Ltd., China - www.safecare.com.cn 

Device identification number
3124
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual
Physical Support
Cassette
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Enterovirus A71 (EV-A71), Hemophilus Influenzae, Human Metapneumovirus (HMPV), MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Rhinovirus, SARS-CoV
Pathogens detected
Influenza A, Influenza B, SARS-CoV
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.525 (Eta), B.1.526 (Iota), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.621 (Mu), C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
No
Assay Type
Immuno-Antigen
Self Test
Yes
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
LOD
130 TCID50/ml
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
0 %
False negatives
13.64 %
Precision
Evaluated
Accuracy
97.35 % (COVID-19)
Accuracy
100 % (Influenza)
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
86.36 % (COVID-19)
Clinical Sensitivity
100 % (Influenza)
Clinical Specificity
100 % (COVID-19)
Clinical Specificity
100 % (Influenza)
Type of antigen
Nucleocapsid protein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements