COVID-19 In Vitro Diagnostic Medical Device - detail

Influenza & COVID-19 Ag Combo Rapid Test

Manufactured by Zhejiang Orient Gene Biotech Co. Ltd., China - https://www.orientgene.com/ 

Device identification number
3095
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual
Physical Support
Cassette
Target type
Antigen
Targets
nucleocapsid protein
Specimen
Nasal swab, Nasopharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Enterovirus A71 (EV-A71), Hemophilus Influenzae, Herpes Simplex (HSV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, Mumps Virus (MuV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus
Pathogens detected
Influenza A, Influenza B, SARS-CoV
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
The Coronavirus Rapid Antigen Test Cassette (swab) from Orient Gene is already on the HSC list. This request is to also register the Influenza & COVID-19 Ag Combo Rapid Test from the same manufacturer.
Assay Type
Immuno-Antigen
Self Test
No
Reader Required
No
Subcategory
Visualisation system
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
LOD
115 TCID50/ml
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
0.77 % (2/258)
False negatives
3.27 % (2/61)
Precision
Evaluated
Accuracy
98.74 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
96.72 %
Clinical Specificity
99.22 %
Type of antigen
Nucleocapsid protein