SARS-CoV-2 Antigen Rapid Test Kit

Manufactured by Xiamen Higia Medical Technology Co. Ltd., China - www.ivdtestkit.com 

Device identification number
3051
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Anterior nasal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Enterovirus A71 (EV-A71), Human Metapneumovirus (HMPV), Influenza A H1N1, Influenza A H3N2, Influenza B Victoria, Influenza B Yamagata, Parainfluenza Virus Type, Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus A, Rhinovirus B
Pathogens detected
Enterovirus A71 (EV-A71), Human Metapneumovirus (HMPV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Rhinovirus A, Rhinovirus B
Information on antigen epitope targeted by the antibodies used in the design of the device
MHHHHHHMSDNGPQNQRNAPRITFGGPSDSTGSNQNGERSGARSKQRRPQGLPNNTASWFTALTQHGKE DLKFPRGQGVPINTNSSPDDQIGYYRRATRRIRGGDGKMKDLSPRWYFYYLGTGPEAGLPYGANKDGIIWVAT EGALNTPKDHIGTRNPANNAAIVLQLPQGTTLPKGFYAEGSRGGSQASSRSSSRSRNSSRNSTPGSSRGTSPAR MAGNGGDAALALLLLDRLNQLESKMSGKGQQQQGQTVTKKSAAEASKKPRQKRTATKAYNVTQAFGRRGPE QTQGNFGDQELIRQGTDYKHWPQIAQFAPSASAFFGMSRIGMEVTPSGTWLTYTGAIKLDDKDPNFKDQVILL NKHIDAYKTFPPTEPKKDKKKKADETQALPQRQKKQQTVTLLPAADLDDFSKQLQQSMSSADSTQA
Lineages detected
B.1.1.529 (Omicron), B.1.617.2 (Delta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
We are registered with the Italian Ministry of Health, and CE certified in the Netherlands
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
20 minutes
Subclass
Sandwich, Double
LOD
2000 TCID50/ml
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
0.22 % (1 false positives out of 450 negative samples)
False negatives
7.8 % (9 false negatives out of 115 positive samples)
Precision
Evaluated
Accuracy
98.2 % (Antigen)
Reproducibility
Not evaluated
Robustness
Not evaluated
Clinical Sensitivity
92.2 % (Antigen)
Clinical Specificity
99.8 (Antigen)
Type of antigen
Nucleocapsid protein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements