InstantSure Covid-19 Ag CARD

Manufactured by Suzhou Soochow University Saier Immuno Biotech Co. Ltd., China

Device identification number
3015
CE Marking
Yes
HSC common list (RAT)
Yes
Format
Near POC / POC
Physical Support
Card
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab, Nasopharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus OC43 (HCoV-OC43), Enterovirus A71 (EV-A71), Human Metapneumovirus (HMPV), Influenza A, Influenza B, MERS-CoV, Mycoplasma Pneumoniae, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV) Type A, Rhinovirus
Pathogens detected
SARS-CoV
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
This kit is used for in vitro qualitative detection of SARS-CoV-2 antigen in human nasopharyngeal swabs, nasal swabs within 7 days after clinical symptoms. This test is intended for use by trained clinical laboratory personnel specifically instructed and trained in the techniques of in vitro diagnostic procedures, and proper infection control procedures and individuals similarly trained in point of care settings.
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Method
Immunoassay
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
LOD
160 TCID50/ml
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
3 AU
False negatives
38 AU
Precision
Evaluated
Accuracy
96.58 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
96.69 %
Clinical Specificity
99.52 %
Type of antigen
Nucleocapsid protein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements