COVID-19 Antigen Rapid Test Device

Manufactured by Hangzhou Aichek Medical Technology Co. Ltd., China - www.aichek.com 

Device identification number
3005
CE Marking
Yes
HSC common list (RAT)
Yes
Format
Manual, Near POC / POC
Physical Support
Cassette, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasopharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Hemophilus Influenzae, Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Rhinovirus, Rhinovirus A, SARS-CoV
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.617.2 (Delta), P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2023-02-01 02:54:39 CET
Comments
Aichek COVID-19 Antigen Rapid Test Device has been listed of BfArm and ANSM.
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
Other ()
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
LOD
50 TCID50/ml
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
% 0 false positives in 350 samples
False negatives
5 % (5 false negatives in 100 samples)
Precision
Evaluated
Accuracy
98.75 % (Antigen)
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95 % (Antigen)
Clinical Specificity
100 % (Antigen)
Type of antigen
Nucleocapsid protein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements