COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Virus Antigen Detection Kit (colloidal gold method)

Manufactured by BGI Europe A/S - http://www.bgi.com 

Device identification number
2988
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Cassette, Lateral flow
Target type
Antigen
Targets
nucleocapsid protein
Specimen
Anterior nasal swab, Nasal swab, Nasopharyngeal swab, Throat swab
Cross-reactivity (pathogens tested)
Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Human Metapneumovirus (HMPV), Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mumps Virus (MuV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV) Type B, Rhinovirus A, Rhinovirus B, SARS-CoV
Pathogens detected
SARS-CoV
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.429 (Epsilon), B.1.525 (Eta), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Assay Type
Immuno-Antigen
Reader Required
No
Method
Immunoassay
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
Detection Principle
Colloidal gold
LOD
200 TCID50/ml
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
4.26 % (4 out of 94 (Ct < 30))
False negatives
0 %
Precision
Evaluated
Accuracy
99.28 % (Ct values < 30)
Accuracy
97.89 % (overall)
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95.74 % (Ct values <30)
Clinical Specificity
100 %
Type of antigen
Nucleocapsid protein