SMARTTEST SARS-COV-2 NASOPHARYNGEAL ANTIGEN RAPID TEST

Manufactured by A PLUS DIAGNOSTICS LAB.SAN.TIC.A.S., Turkey - www.aplusıvd.com 

Device identification number
2985
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual
Physical Support
Cassette, Lateral flow, Strip
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab, Nasopharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus 3, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Epstein-Barr Virus (EBV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 2, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus
Pathogens detected
SARS-CoV
Lineages detected
B.1.1.529 (Omicron), B.1.351 (Beta), B.1.617.2 (Delta), P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
The test was approved by BFARM
Assay Type
Immuno-Antigen
Self Test
Yes
Subcategory
Other ()
Method
Immunoassay
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
LOD
100 TCID50/ml
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
1.99 % (9 out of 452 samples were found falsely positive comparing with the PCR test)
False negatives
0.95 % (5 out of 523 samples were found falsely negative comparing with the PCR test)
Precision
Evaluated
Accuracy
98.3 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
98.26 %
Clinical Specificity
99.05 %
Type of antigen
Nucleocapsid protein