COVID-19 Antigen Rapid Test

Manufactured by Changchun WanCheng Bio Electron Co. Ltd., China - https://jlwancheng.en.alibaba.com 

Device identification number
2934
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual, Near POC / POC
Physical Support
Cassette, Lateral flow, Strip
Target type
Antigen
Targets
nucleocapsid protein
Specimen
Nasal swab, Nasopharyngeal swab, Oropharyngeal swab, Saliva, Throat swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Anti-Nuclear Antibody, Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Cytomegalovirus (CMV), Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Hemophilus Influenzae, Hepatitis A Virus (HAV), Hepatitis B Virus (HBV), Hepatitis C Virus (HCV), Herpes Simplex (HSV), Human Immunodeficiency Virus (HIV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mumps Virus (MuV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV, Varicella Zoster Virus (VZV)
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
None
Assay Type
Immuno-Antigen
Self Test
Yes
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
LOD
100 TCID50/ml
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
0.7 % (No more than 0.7%)
False negatives
1.5 % (No more than 1.5%)
Precision
Evaluated
Accuracy
99.6 % (Antigen)
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
98.8 % (Antigen)
Clinical Specificity
100 % (Antigen)
Type of antigen
Nucleocapsid protein