Skip to main content
COVID-19 In Vitro Diagnostic Devices and Test Methods Database

COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 (S-RBD & NP) Antibody Test (Colloidal Gold)

Manufactured by Beijing Hotgen Biotech Co. Ltd., China - www.hotgen.com.cn 

Device identification number
2811
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Cassette
Target type
Antibody
Specimen
Plasma, Serum, Venous whole blood, Whole blood with anti-coagulants
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Coronaviruses (HCoV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV
Lineages detected
B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, C.37 (Lambda), P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
Two test results can be obtained through this product test: S-RBD antibody and NP antibody.
Assay Type
Immuno-Antibody
Reader Required
No
Subcategory
Other ()
Method
Immunochromatography
Measurement
Semiquantitative
Time
20 minutes
Subclass
Sandwich
Detection Principle
Colloidal gold
LOD
15 AU/mL (15BAU/mL)
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
0 % 408 negative samples have no false positives.
False negatives
0.43 % (230 positive samples have 1 false negative.)
Precision
Evaluated
Accuracy
99.84 % IgG
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
99.57 % IgG
Clinical Specificity
100 % IgG

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements