COVID-19 In Vitro Diagnostic Medical Device - detail

COVID-19 TEST RAPIDO ANTIGENE RICOV4

Manufactured by Beijing North Institute of Biotechnology Co. Ltd., China - www.bnibt.com 

Device identification number
2733
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Cassette, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Influenza A, Influenza A H1N1, Influenza B, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Respiratory Syncytial V (RSV), Rhinovirus A, SARS-CoV
Information on antigen epitope targeted by the antibodies used in the design of the device
clone1 binding site:INTNSSP(74-80) clone2 binding site:GTRNPANN(147-154)
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.617.1 (Kappa), B.1.617.2 (Delta), C.37 (Lambda)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
2022.2.7 evaluated at Centralne Laboratorium Kliniczne Uniwersyteckie Centrum Kliniczne.
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Reader Required
No
Subcategory
Visualisation system
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
LOD
150 TCID50/ml
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
0.7 %
False negatives
9.3 %
Precision
Evaluated
Accuracy
97.67 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
90.74 %
Clinical Specificity
99.33 %
Type of antigen
Nucleocapsid protein