Novel Coronavirus (COVID-19) Antigen Test Kit (Colloidal Gold) - (saliva, sputum)

Manufactured by Hangzhou Laihe Biotech Co. Ltd., China - www.lyherbio.com 

Device identification number
2634
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual, Near POC / POC
Physical Support
Cassette
Target type
Antigen
Targets
nucleocapsid protein
Specimen
Saliva, Sputum
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Anti-Nuclear Antibody, Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Cytomegalovirus (CMV), Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Hemophilus Influenzae, Hepatitis A Virus (HAV), Hepatitis B Virus (HBV), Hepatitis C Virus (HCV), Herpes Simplex (HSV), Human Immunodeficiency Virus (HIV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mumps Virus (MuV), Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV, Varicella Zoster Virus (VZV)
Lineages detected
A.23.1, AT.1, B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
lineages performances: Omicron (B.1.1.529), Mu (B.1.621), Lambda (C.37), Delta (B.1.617.2), Alpha (B.1.1.7), Beta (B.1.351, Gamma (P.1)
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Method
Immunoassay
Measurement
Qualitative
Time
15 minutes
Detection Principle
Colloidal gold
LOD
0.5 ng/mL
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
0.52 (Tested out 1 false positive result from 189 negative cases.)
False negatives
4.42 (Tested out 5 false negative results from 113 positive cases.)
Precision
Evaluated
Accuracy
98.01 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95.58 %
Clinical Specificity
99.47 %
Type of antigen
Nucleocapsid protein