Skip to main content
COVID-19 In Vitro Diagnostic Devices and Test Methods Database

COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Antigen Rapid Test Cassette

Manufactured by Merlin Biomedical (Xiamen) Co. Ltd., China - http://www.merlinbio.com.cn/en/ 

Device identification number
2596
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual
Physical Support
Cassette
Target type
Antibody
Targets
nucleocapsid protein
Specimen
Saliva
Cross-reactivity (pathogens tested)
SARS-CoV
Lineages detected
A.23.1, AT.1, B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
This cassette belongs to in vitro diagnostic reagents, and is suitable for the qualitative cietection cf SARS-CoV-2's Nucleocapsid protein in human saliva specimen. It cannot be served as the basis of the confirmation or exclusion for coronavirus pneumonia. This product is used only far medicai institutions. The positive test result requires further confirmation. A negative test cannot rule out the possibility of infection. The test results of this cassette are only for clinica! reference. It is recommended to conduct a comprehensive analysis of the patient's condition in combination with clinical manifestations and other laboratory tests. A total of 443 clinical samples were tested in this clinical study. The clear diagnosis and exclusion criteria of the "SARS-CoV-2 diagnosis and treatment protocol" were taken as a reference. Clinical trials were conducted at Fujian's Fuzhou Lung Hospital and Zhangzliou Hospital. The positive coincidence rate of the SARS-CoV-2 Antigen Rapid Test Cassette, developed by Merlin was 95.5% and the negative coincidence rate of the cassette was 99.7%, the total coincidence rate of the cassette was 98 , 4% compared to the reference method (RT-PCR) which indicates a high consistency.
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Method
Immunochromatography
Time
15 minutes
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
4.5 % (Positive coincidence rate = 127/ (127+6) x100%=95.5%, (95% Cl: 90.5%-98.0%))
False negatives
0.3 % (Negative coincidence rate = 299/ (299+1) x100%=99.7%, (95% Cl: 98.1%-99.9%);)
Precision
Evaluated
Accuracy
98 % ( (Antigen))
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95 % ( (Antigen))
Clinical Specificity
99 % ( (Antigen))
Type of antigen
Nucleocapsid protein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements