COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Rapid Antigen Test 2.0

Manufactured by Roche (SD BIOSENSOR), Switzerland - https://diagnostics.roche.com/ 

Device identification number
25546
CE Marking
Yes
HSC common list (RAT)
Yes
Format
Near POC / POC
Physical Support
Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasopharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 7, Adenovirus Subtype B, Adenovirus Subtype C, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Coronaviruses (HCoV), Hemophilus Influenzae, Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mycoplasma Pneumoniae, Other (Enterovirus type 68, Streptococcus pneumonia, Streptococcus pyrogenes, Candida albicans, Legionella pneumophila, Staphylococcus aureus, Staphylococcus epidermidis, Pooled human nasal wash), Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV, SARS-CoV-2
Pathogens detected
SARS-CoV, SARS-CoV-2
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), Other (), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2023-03-31 05:45:26 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
N.A.
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
Detection Principle
Colloidal gold
LOD
14.6 TCID50/ml
Calibration
Not evaluated
Analytical Sensitivity
100 % (Refers to the LoD and analytical studies)
Analytical Specificity
100 % (Refers to the analytical studies)
Analysis of cross reactivity
Evaluated
False positives
1 tests (1 out of 402)
False negatives
1 tests (1 out of 100)
Precision
Evaluated
Accuracy
100 % (Not applicable for rapid tests)
Robustness
Evaluated
Clinical Sensitivity
99 % (Overall Ct values)
Clinical Specificity
99.75 % (Overall Ct values)

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements