COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Antigen Rapid Test (Nasal Swab) for self testing

Manufactured by Hangzhou AllTest Biotech Co. Ltd., China - www.alltests.com.cn 

Device identification number
2529
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Cassette, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasal swab
Cross-reactivity (pathogens tested)
Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Influenza A H1N1, Influenza A H3N2, Influenza B, MERS-CoV, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Respiratory Syncytial V (RSV)
Pathogens detected
Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Influenza A H1N1, Influenza A H3N2, Influenza B, MERS-CoV, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Respiratory Syncytial V (RSV)
Type of Antigen (in case of Antigen-based)
Nucleocapsid protein
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-10-12 09:39:01 CET
Comments
The SARS-CoV-2Antigen Rapid Test(Nasal Swab)is a single-use test kit intended to detect the SARS-CoV-2 that causes COVID-19 with self-collected nasal swab specimen from symptomatic individuals who are suspected of being infected with COVID-19. Results are for the detection of SARS-CoV-2 Nucleocapsid protein Antigens.An antigen is generally detectable in upper respiratory specimens during the acute phase of infection. Positive results indicate the presence of viral antigens, but clinical correlation with patient history and other diagnostic information is necessary to determine infection status. Positive resultsare indicative of thepresence of SARS-CoV-2. ndividuals who test positive should self-isolate and seek additional care from their healthcare provider. Positive results do not rule out bacteria infection or co-infection with other viruses. Negative results do not precludeSARS-CoV-2 infection. Individuals who test negative and continue to experience COVID-like symptoms should seek follow up care from their healthcare provider.
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
Yes
Reader Required
No
Subcategory
Validity test kit
Method
Immunoassay
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich
Detection Principle
Immunofluorescence
LOD
100 AU
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
0.028 %
False negatives
0.005 %
Precision
Evaluated
Accuracy
98.6 % (Antigen)
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
97.2 % (Antigen)
Clinical Specificity
99.5 % (Antigen)
Type of antigen
Nucleocapsid protein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements