Orawell COVID-19 Ag Rapid Saliva Test Device (CO-03)
Manufactured by Jiangsu Well Biotech Co. Ltd., China - www.wellbioscience.com
Device identification number
2500
CE Marking
✓Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Cassette
Target type
Antigen
Specimen
Saliva
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Coronaviruses (HCoV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza B, MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Rhinovirus, SARS-CoV
Lineages detected
B.1.1.7 (Alpha), B.1.351 (Beta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Assay Type
Immuno-Antigen
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
10 minutes
Subclass
Sandwich, Double
LOD
60 TCID50/ml
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False negatives
4.4 %
Precision
Evaluated
Accuracy
98.21 % ((Antigen))
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95.6 % ((Antigen))
Clinical Specificity
100 % ((Antigen))
Type of antigen
Nucleocapsid protein