Skip to main content
COVID-19 In Vitro Diagnostic Devices and Test Methods Database

COVID-19 In Vitro Diagnostic Medical Device - detail

FaStep COVID-19 Antigen Nasal Test Kit

Manufactured by Assure Tech (Hangzhou) Co. Ltd., China - www.diareagent.com 

Device identification number
2469
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Lateral flow
Target type
Antigen
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Chlamydia Pneumoniae, Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Hemophilus Influenzae, Human Metapneumovirus (HMPV), Influenza A H1N1, Influenza A H3N2, Influenza B Victoria, Influenza B Yamagata, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus A, Rhinovirus B, SARS-CoV
Lineages detected
B.1.1.7 (Alpha), B.1.351 (Beta), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
None
Assay Type
Immuno-Antigen
Reader Required
No
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
LOD
251 AU
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
0 % none, 100% specific
False negatives
0 % no false negatives occurred on antigen level at 370 pg/mL or virus level at 1×10 2.4 TCID50/mL
Precision
Evaluated
Accuracy
98.7 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
96.1 %
Clinical Specificity
99.8 %
Type of antigen
Nucleoprotein