COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Antigen Rapid Test (Immunochromatography)

Manufactured by Qingdao Hightop Biotech Co. Ltd., China - www.hightopbio.com 

Device identification number
2191
CE Marking
Yes
HSC common list (RAT)
×No
Format
Manual
Physical Support
Cassette
Target type
Antigen
Specimen
Nasal swab, Nasopharyngeal swab
Cross-reactivity (pathogens tested)
SARS-CoV
Lineages detected
A.23.1, AT.1, B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Method
Immunochromatography
Time
15 minutes
LOD
2000 AU
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
5 AU
False negatives
0.98 % (1/102)
Precision
Evaluated
Accuracy
98 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95 %
Clinical Specificity
99 %
Type of antigen
Nucleoprotein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements