COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 Antigen Rapid Test (Lateral Flow Assay)

Manufactured by Shenzhen Kang Sheng Bao Bio-Technology Co. Ltd., China

Device identification number
2166
CE Marking
Yes
HSC common list (RAT)
×No
Format
Lab-based, Manual, Near POC / POC
Physical Support
Cassette
Target type
Antigen
Targets
nucleocapsid protein
Specimen
Anterior nasal swab, Nasopharyngeal swab, Oropharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus 7, Chlamydia Pneumoniae, Epstein-Barr Virus (EBV), Influenza A H1N1, Influenza A H3N2, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Mumps Virus (MuV), Mycoplasma Pneumoniae, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B
Lineages detected
B.1.1.7 (Alpha), B.1.351 (Beta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
no additional information
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Reader Required
No
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
LOD
0.05 AU
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
1 %
False negatives
4.81 %
Precision
Evaluated
Accuracy
96.71 % ((Antigen))
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95.19 % ((Antigen))
Clinical Specificity
98.17 % ((Antigen))
Type of antigen
Nucleoprotein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements