SARS CoV 2 Ag Saliva Rapid Test
Manufactured by VivaChek Biotech (Hangzhou) Co. Ltd., China - www.vivachek.com
Device identification number
2118
CE Marking
✓Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Lateral flow
Target type
Antigen
Specimen
Saliva
Cross-reactivity (pathogens tested)
Adenovirus 3, Adenovirus 7, Adenovirus Subtype B, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), MERS-CoV, Mycobacterium Tuberculosis, Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4
Lineages detected
B.1.1.7 (Alpha), B.1.351 (Beta), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
SARS-CoV-2 Ag Saliva Rapid Test is for the rapid, qualitative detection of the nucleocapsid protein antigen from SARS-CoV-2 in human Saliva
Assay Type
Immuno-Antigen
Reader Required
No
Subcategory
Validity test kit
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich
LOD
75.5 AU
Calibration
Not evaluated
Analysis of cross reactivity
Evaluated
False positives
0 % 431 negative Sample were tested and 0 of them were false positive
False negatives
1.74 % (172 positive Sample were tested and 3 of them were false negative.)
Precision
Evaluated
Accuracy
99.5 %
Reproducibility
Evaluated
Robustness
Not evaluated
Clinical Sensitivity
98.26 %
Clinical Specificity
99.9 %
Type of antigen
Nucleoprotein