COVID-19 In Vitro Diagnostic Medical Device - detail

SARS-CoV-2 antigen IVD kit SWAB

Manufactured by Shenzhen Reagent Technology Co. Ltd., China - https://www.reagen.cn/ 

Device identification number
2026
CE Marking
Yes
HSC common list (RAT)
Yes
Format
Manual, Near POC / POC
Physical Support
Card, Lateral flow
Target type
Antigen
Antibody type
Monoclonal
Targets
nucleocapsid protein
Specimen
Nasopharyngeal swab, Oropharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Coronaviruses (HCoV), Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, Mumps Virus (MuV), Mycoplasma Pneumoniae, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Respiratory Syncytial V (RSV) Type A, Respiratory Syncytial V (RSV) Type B, Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV
Lineages detected
B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, C.37 (Lambda), P.1 (Gamma), P.2 (Zeta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
Dear Team, We would like submit additional information about the mutated varaints. eg. Omicron and Delta. Thank you for your support! For further questions, please feel free to contact us. Best regards, Team of Shenzhen Reagent Co., Ltd.
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Self Test
No
Reader Required
No
Subcategory
Other ()
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Membrane-based
LOD
200 TCID50/ml
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
1.9 % (2 false positives were detected in 155 negative samples)
False negatives
4.76 % (5 false negatives were detected in 105 positive samples)
Precision
Evaluated
Accuracy
97.31 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
95.23 % ((95%CI:85.94%-98.86%))
Clinical Specificity
98.71 % ((95%CI:86.75%-99.12%))
Type of antigen
Nucleocapsid protein

The database contains publicly available In Vitro Diagnostic Medical Devices for COVID-19 and it is being updated periodically. Please note that additional performance (as retrieved from manufacturers web pages) is provided only for devices commercially available with CE-IVD mark. Acknowledgements