SieloCovidAg
Manufactured by DyonMed SA, Greece - www.dyonmedical.com
Device identification number
2007
CE Marking
✓Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Cassette
Target type
Antigen
Specimen
Nasal swab, Nasopharyngeal swab, Oropharyngeal swab, Saliva, Sputum
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Bordetella Pertussis, Coronaviruses (HCoV), Cytomegalovirus (CMV), Enterovirus A71 (EV-A71), Herpes Simplex (HSV), Influenza A, Influenza B, MERS-CoV, Mumps Virus (MuV), Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Respiratory Syncytial V (RSV), Rhinovirus, Rhinovirus A
Lineages detected
A.23.1, AT.1, B.1.1.529 (Omicron), B.1.1.7 (Alpha), B.1.351 (Beta), B.1.427 (Epsilon), B.1.429 (Epsilon), B.1.525 (Eta), B.1.526 (Iota), B.1.526.1, B.1.526.2, B.1.616, B.1.617.1 (Kappa), B.1.617.2 (Delta), B.1.617.3, B.1.621 (Mu), C.36, C.37 (Lambda), P.1 (Gamma), P.3 (Theta)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
NO OTHER
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Reader Required
No
Method
Immunochromatography
Measurement
Qualitative
Time
11 minutes
Subclass
Sandwich, Double
LOD
6 AU
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
0 AU
False negatives
6 AU
Precision
Evaluated
Accuracy
98 % ((Antigen))
Clinical Sensitivity
95.6 % ((Antigen))
Clinical Specificity
99 % ((Antigen))
Type of antigen
Nucleocapsid protein