Skip to main content
COVID-19 In Vitro Diagnostic Devices and Test Methods Database

COVID-19 In Vitro Diagnostic Medical Device - detail

CORA CHECK-19

Manufactured by fameditec GmbH, Germany - https://www.fameditec.de/ 

Device identification number
1744
CE Marking
Yes
HSC common list (RAT)
×No
Format
Near POC / POC
Physical Support
Cassette
Target type
Antigen
Specimen
Anterior nasal swab, Nasal swab, Nasopharyngeal swab, Oropharyngeal swab
Cross-reactivity (pathogens tested)
Adenovirus, Adenovirus 3, Adenovirus 7, Alpha Coronavirus 229E (HCoV-229E), Alpha Coronavirus Nl63 (HCoV-Nl63), Beta Coronavirus HKU1 (HCoV-HKU1), Beta Coronavirus OC43 (HCoV-OC43), Coronaviruses (HCoV), Enterovirus A71 (EV-A71), Epstein-Barr Virus (EBV), Human Metapneumovirus (HMPV), Influenza A, Influenza A H1N1, Influenza A H3N2, Influenza A H5N1, Influenza B, Influenza B Victoria, Influenza B Yamagata, MERS-CoV, Parainfluenza Virus Type, Parainfluenza Virus Type 1, Parainfluenza Virus Type 2, Parainfluenza Virus Type 3, Parainfluenza Virus Type 4, Respiratory Syncytial V (RSV), Rhinovirus, Rhinovirus A, Rhinovirus B, SARS-CoV
Lineages detected
B.1.1.7 (Alpha), B.1.351 (Beta), P.1 (Gamma)
Commercial Status
Commercialised
Last Update
2022-08-24 09:08:01 CET
Comments
https://www.fameditec.de/
Assay Type
Immuno-Antigen
Rapid Diagnostic
Yes
Reader Required
No
Method
Immunochromatography
Measurement
Qualitative
Time
15 minutes
Subclass
Sandwich, Double
Detection Principle
Colorimetry
LOD
175 AU
Calibration
Evaluated
Analysis of cross reactivity
Evaluated
False positives
0.72 %
False negatives
3.51 %
Precision
Evaluated
Accuracy
98.47 %
Reproducibility
Evaluated
Robustness
Evaluated
Clinical Sensitivity
96.49 %
Clinical Specificity
98.28 %
Type of antigen
Nucleocapsid protein