Diagnostic accuracy of an automated chemiluminescent immunoassay for anti-SARS-CoV-2 IgM and IgG antibodies: an Italian experience
Detection Principle
ImmunoAssay-Antibody
Target
IgM, IgG
Testing Method Category
Testing Method
kit
Testing Method - Additional Info
iFlash1800 CLIA analyzer from YHLO biotechnology Co (LTD, Shenzhen, China)
Reported Performance
sensitivity: 73.3%(IgM), 76.7% (IgG); specificity: 92.2% (IgM), 100% (IgG). Remark: results improve modifying cutoff
Sample Size
61 patients & 64 controls
Peer-reviewed
yes
The database contains available information from scientific literature that is being updated periodically. Please note that the provided information (as retrieved from analysed papers) is provided only for devices commercially available with CE-IVD mark. Acknowledgements