Diagnostic accuracy of an automated chemiluminescent immunoassay for anti-SARS-CoV-2 IgM and IgG antibodies: an Italian experience

Detection Principle
ImmunoAssay-Antibody
Target
IgM, IgG
Testing Method Category
 
Testing Method
kit
Testing Method - Additional Info
iFlash1800 CLIA analyzer from YHLO biotechnology Co (LTD, Shenzhen, China)
Reported Performance
sensitivity: 73.3%(IgM), 76.7% (IgG); specificity: 92.2% (IgM), 100% (IgG). Remark: results improve modifying cutoff
Sample Size
61 patients & 64 controls