High-throughput immunoassays for SARS-CoV-2 – considerable differences in performance when comparing three methods

Detection Principle
ImmunoAssay-Antibody
Target
IgM and IgG
Testing Method Category
Other
Testing Method
1) Abbott SARS-CoV-2 IgG on the ARCHITECT i2000 (Abbott, Illinois, USA); 2) Elecsys Anti-SARS-CoV-2 on the Cobas 8000 e801 (Roche Diagnostic Scandinavia AB, 80 Solna, Sweden); 3) LIAISON SARS-CoV-2 S1/S2 IgG on the LIAISON XL (DiaSorin, 81 Saluggia, Italy); 4) the lateral flow test 2019-nCOV IgG/IgM Rapid Test (Dynamiker 82 Biotechnology Co., Tianjin, China)
Testing Method - Additional Info
Performances opf three different comercially available high-throughput antibody assay and one rapid immunochromatographic test
Reported Performance
1) Abbott Sensitivity: 85%; Specificity: 100%; 2) Roche Sensitivity: 100%; Specificity: 98% 3) DiaSorin Sensitivity: 77.8%; Specificity: 96.9%; 4) Dynamiker Sensitivity: 75%; Specificity: 98%
Sample Size
133 patients
Peer-reviewed
no